Recall record at a glance
| Official record | Z-0207-2016 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Arrow International Inc. |
| Report date | Nov 11, 2015 |
| Recall initiation date | Oct 16, 2015 |
| Quantity | 12,157 |
| Location | Reading, PA, United States |
Official product description
ARROW PICC Kit featuring Chlorag+ard Technology Midline Catheterization Kit ARROW Advantage5 Pressure Injectable PICC Kit Arrow Pressure Injectable Arrow PICC powered by Arrow VPS Stylet ARROW Pressure Injectable JACC with Chlorag+ard Technology For short-term or long-term peripheral access to the central venous system for intravenous therapy, blood sampling, infusion, pressure injection of contrast media, and allows for central venous pressure monitoring. Permits venous access to the peripheral circulation. Offers an alternative method of intravenous access for select adult and pediatric patients. The ARROW GlideThru Peel-Away Sheath/Dilator is used for the percutaneous introduction of diagnostic or therapeutic devices into the vasculature.
Why the product was recalled
Possibility that the catheter peel-away component hub tabs may prematurely detach when the practitioner begins to peel apart the sheath body from the catheter. Although the issue does not pose a significant risk to the patient, it may cause a delay of treatment when placing the catheter.
Distribution information
World Wide Distribution - US including AR, AL, TX, MA, CA, IA, MO, NY, AL, VA, RI, IN, TN, FL, KY, NH, SC, MD, WA, OK, IL, ME, OR, OH, PA, MI, KS, LA, TN, WV, NE, NJ, NV, GA, SD, CT, MS, CO, and Internationally to Canada and Belgium.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.