FDA recall record

Arrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT2 WAVE JAPANESE, Product Code IAP-0500J recall — FDA Z-2413-2020

A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is…

Class Iclassification
Z-2413-2020official record ID
Jul 8, 2020report date

Recall record at a glance

Official recordZ-2413-2020
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusTerminated
Recalling firmArrow International Inc.
Report dateJul 8, 2020
Recall initiation dateMay 20, 2020
Quantity2123 total
LocationReading, PA, United States

Official product description

Arrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT2 WAVE JAPANESE, Product Code IAP-0500J

Why the product was recalled

A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector wires, which may result in pump alarms for System Error 3 and High Baseline presented on the screen of the IABP, and potential abrupt cessation of function or inability to start the IABP.

Distribution information

Worldwide distribution.

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.