Recall record at a glance
| Official record | Z-2733-2014 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Lumenis, Inc. |
| Report date | Oct 8, 2014 |
| Recall initiation date | Aug 15, 2014 |
| Quantity | 16 |
| Location | Salt Lake City, UT, United States |
Official product description
Array LaserLink, Manufactured by Lumenis, The Array" LaserLink" is a laser system accessory intended for use in the treatment of ocular pathology. " For the Posterior Segment, the Array" LaserLink" is indicated for use in Retinal Photocoagulation and Panretinal Photocoagulation of Vascular and Structural Abnormalities of the Retina and Choroid including: " Proliferative and Severe and Very Severe Non-Proliferative Diabetic Retinopathy " Macular Edema associated with Proliferative or Non-Proliferative Diabetic Retinopathy " Choroidal Neovascularization " Retinal Neovascularization associated with Retinal Occlusive Disease (Branch Retinal Vein Occlusion; Central Retinal Vein Occlusion) " Macular Edema associated with Branch Retinal Vein Occlusion " Retinal Tears and Detachments And Anterior Segments as follows: " Iridotomy in Closed Angle Glaucoma " Trabeculoplasty in Open Angle Glaucoma
Why the product was recalled
Lumenis initiated a field-correction for the Array Laser Link", GA-0006700 (SN XXYYZZ) because the system may project an erroneous pattern display on the retina, which is different than the desired pattern.
Distribution information
Worldwide Distribution - USA (nationwide) and Internationally to JAPAN, CHINA, and CANADA.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Visit.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.