FDA recall record

Arjo Hospital Equipment AB ArjoHuntleigh Contoura C1000 series beds are designed to recall — FDA Z-0019-2015

There is a low but existing entrapment risk for a patient being treated on the Contoura 1000 and Contoura 1080 bariatric beds if they have a BMI < 40.

Class IIclassification
Z-0019-2015official record ID
Oct 15, 2014report date

Recall record at a glance

Official recordZ-0019-2015
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmArjo Hospital Equipment AB
Report dateOct 15, 2014
Recall initiation dateMar 17, 2014
Quantity1,695 beds (worldwide)
LocationEslov, Sweden

Official product description

ArjoHuntleigh Contoura C1000 series beds are designed to accommodate patients weighing up to 990 lbs. (450 kg). Model C1000 is the basic bed. Model C1080 includes a weighing system and a power outlet for an ArjoHuntleigh approved mattress pump. Product Usage: The Contoura series bariatric beds are electrically operating medical beds with a maximum patient weight of 450 kg. This range is intended for use in hospitals to facilitate management of obese patients. An optional integrated weighing system helps with assessment and management of the patient if required.

Why the product was recalled

There is a low but existing entrapment risk for a patient being treated on the Contoura 1000 and Contoura 1080 bariatric beds if they have a BMI < 40.

Distribution information

Worldwide Distribution - US: Nationwide in the states of (AZ, CA, CO, CT, GA, ID, IL, MA, MD, MN, NC, NE, OH, PA, RI, SD, TX, VA, WI); and the countries of Australia, Austria, Bahrain, Belgium, Canada, Denmark, France, Germany, Greece, Iceland, India, Ireland, Israel, Italy, Lebanon, Mexico, Netherlands, New Zealand, Norway, Saudi Arabia, Singapore, Spain, Sweden, Switzerland, United Arab Emirates

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.