FDA recall record

Securitas Healthcare Arial mobile application, model numbers 54640, 54630, and 54630-SMA recall — FDA Z-1943-2024

The Arial Mobile Application utilizing Apple iOS 17.4 and iOS 17.5 is frequently restarting causing the app to receive alerts for all pending alarm…

Class IIclassification
Z-1943-2024official record ID
Jun 12, 2024report date

Recall record at a glance

Official recordZ-1943-2024
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmSecuritas Healthcare LLC
Report dateJun 12, 2024
Recall initiation dateApr 11, 2024
Quantity661 sites with mobile application user licenses
LocationLincoln, NE, United States

Official product description

Arial mobile application, model numbers 54640, 54630, and 54630-SMA, that utilizes Apple iOS 17.4 and Apple iOS 17.5, a component of the Arial Emergency and Nurse Call System.

Why the product was recalled

The Arial Mobile Application utilizing Apple iOS 17.4 and iOS 17.5 is frequently restarting causing the app to receive alerts for all pending alarm notifications and other events when the app restarts, experience difficulty in placing and maintaining staff-to-staff phone calls, and frequently appears offline while the app restarts.

Distribution information

Worldwide distribution - US Nationwide and the countries of Australia, Canada, England, Hong Kong, France, Qatar, and Spain.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records