FDA recall record

Medivance Arctic Sun Temperature Management Systems, Name/REF: Arctic Sun STAT recall — FDA Z-2457-2024

If temperature management system fails to reach correct target water temperature while device is operating in patient control mode, due to sudden patient…

Class IIclassification
Z-2457-2024official record ID
Aug 7, 2024report date

Recall record at a glance

Official recordZ-2457-2024
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmMedivance Inc.
Report dateAug 7, 2024
Recall initiation dateJun 27, 2024
Quantity10507
LocationLouisville, CO, United States

Official product description

Arctic Sun Temperature Management Systems, Name/REF: Arctic Sun STAT, NA/6000-00-00; Arctic Sun STAT Loaner, NA/6000-00-00L; Arctic Sun Stat EU UK/600001; Arctic Sun Stat Taiwan/600006; Arctic Sun Stat Japan/600020; Arctic Sun 5000, NA/5000-00-00; Arctic Sun 5000, NA w/ Transmission Interface Module/5000-00-00E; Loaner, Arctic Sun M5000, NA/5000-00-00L; Arctic Sun 5000, EU/5000-1-01; Loaner, Arctic Sun M5000, EU/5000-01-01L; Arctic Sun 5000, UK/5000-01-02; Arctic Sun 5000, Australia/5000-01-03; Arctic Sun 5000, China/5000-01-04; Arctic Sun 5000, Brazil/5000-01-05; Arctic Sun 5000, Switzerland/5000-01-07; Arctic Sun 5000, South Africa/5000-01-08

Why the product was recalled

If temperature management system fails to reach correct target water temperature while device is operating in patient control mode, due to sudden patient temperature changes/interruption in water flow/blockage of air flow by an obstruction or dirty filter, then system may not alert, and alarm absence may lead to hypothermia or hyperthermia, so a software update will be released to correct issue.

Distribution information

Worldwide - US Nationwide distribution including in the states of IN, OK, MO, FL, CT, KS, TX, NJ, CA, OR, HI, IL, WI, VA, KY, NH, CO, NV, LA, NC, ID, UT, GA, SC, TN, MD, PA, IA, MA, OH, NY, DE, MI, NM, WV, AZ, DC, VT, MN, SD, AL, AR, WA, NE, ND, MT, AK, MS, ME, WY, RI, PR, OK, OR and the countries of CA, DE, AT, BE, FI, TR, NO, DK, JP, ES, AU, FR, GB, IT, SE, PT, AE, SA, ZA, RU, KR, CY, TH, GR, TW, MY, OM, CN, SG, GI, NL, IL, CZ, RO, CH, RE, PL, PE, SK, AR, IN, CL, MX, QA, BR, SI, HU, JO, LU, KW, PH, NZ, HK, VN, IR, BN, UA, ID, CO, KZ, HR, IS, GU, GE, BH, EE, AM, PA, KR, CL, TH, MY, JP, AU, BD, BE, TW

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.