FDA recall record

PROCEPT BIOROBOTICS AquaBeam Handpiece, Sterile EO, Rx Only, REF: 310301, and 320301, a recall — FDA Z-1110-2019

Certain lots may develop an internal leak due to a micro-crack in a specific weld joint within the device. When the Robot detects this failure, it…

Class IIclassification
Z-1110-2019official record ID
Apr 17, 2019report date

Recall record at a glance

Official recordZ-1110-2019
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmPROCEPT BIOROBOTICS CORPORATION
Report dateApr 17, 2019
Recall initiation dateMar 6, 2019
Quantity214
LocationRedwood City, CA, United States

Official product description

AquaBeam Handpiece, Sterile EO, Rx Only, REF: 310301, and 320301, a component of AquaBeam Robotic System. 320301 distributed outside the U.S. Device Description The AQUABEAM Robotic System is intended for use in patients suffering from lower urinary tract symptoms resulting from benign prostatic hyperplasia (BPH).

Why the product was recalled

Certain lots may develop an internal leak due to a micro-crack in a specific weld joint within the device. When the Robot detects this failure, it immediately renders the device inoperable and notifies the user via an error message, which may cause delay and/or inability to complete the procedure.

Distribution information

Worldwide Distribution: US (nationwide) state to: TX, FL, IN, CA, IL, VA, MI, PA.; and (OUS) countries of: Italy, Germany, France, Poland, United Kingdom, Spain, Canada and Australia.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.