FDA recall record

The Anspach Effort Anspach Effort, Inc., 98-0061 MA-15C Bearing Sleeve with Removable Bur recall — FDA Z-0783-2013

Contact between the bur and bur guard could generate metal fragments that may or may not be visible to the surgeon and can potentially remain in the…

Class IIclassification
Z-0783-2013official record ID
Feb 13, 2013report date

Recall record at a glance

Official recordZ-0783-2013
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmThe Anspach Effort, Inc.
Report dateFeb 13, 2013
Recall initiation dateSep 5, 2012
Quantity8 devices
LocationPalm Beach Gardens, FL, United States

Official product description

Anspach Effort, Inc., 98-0061 MA-15C Bearing Sleeve with Removable Bur Guard. A reusable device used with dissection tools; designed for Transphenoidal and Skull base procedures.

Why the product was recalled

Contact between the bur and bur guard could generate metal fragments that may or may not be visible to the surgeon and can potentially remain in the surgical site. The materials used to fabricate the bur guard are not traceable, design validation did not effectively evaluate adequate protection of adjacent tissue and inspection results for each of these bur guards were not documented.

Distribution information

Distributed in the state of Massachusetts

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.