FDA recall record

American Medical Systems AMS 700 Series Inflatable Penile Prosthesis Accessory Kit recall — FDA Z-0209-2015

AMS' supplier has issued a recall of certain lots of Lightning Bolt Keith needles supplied in AMS 700 Accessory Kits. Based on this, AMS is recalling AMS…

Class IIclassification
Z-0209-2015official record ID
Nov 19, 2014report date

Recall record at a glance

Official recordZ-0209-2015
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmAmerican Medical Systems, Inc.
Report dateNov 19, 2014
Recall initiation dateOct 10, 2014
Quantity184 (110 US, 74 OUS)
LocationMinnetonka, MN, United States

Official product description

AMS 700 Series Inflatable Penile Prosthesis Accessory Kit (AMS 700 Line Accessory Kit), 72401850. Keith Needles are part of the AMS 700 Accessory Kit. In conjunction with the Furlow Insertion Tool (Furlow), the Keith Needles facilitate the passing of the pulling sutures through the glans to place the AM3-700 cylinders in the corpus cavernosa.

Why the product was recalled

AMS' supplier has issued a recall of certain lots of Lightning Bolt Keith needles supplied in AMS 700 Accessory Kits. Based on this, AMS is recalling AMS 700 Accessory Kits which contain these lots of needles. The supplier made the decision to issue this recall after an internal investigation that was requested by AMS. The investigation identified a potential nonconformity that may lead to a bent or broken Keith Needle when used with the Furlow tool.

Distribution information

Worldwide Distribution US (nationwide) including PR and the states of AL. AZ. AR, CA, CO, CT, IL, IN, KS, KY, MD, MA,MI, MN, MO, NY,NC, OH,OR, TN, TX, VA, and WA, and the countries of ARGENTINA, AUSTRALIA, BRAZIL, CANDA, GERMANY, FRANCE, GREAT BRITAIN, LEBANON, PANAMA, SOLOMON ISLANDS, SWEDEN, and COLOMBIA.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.