FDA recall record

Abbott Amplatzer TorqVue LP Delivery System (TVLP), Model nos. 9-TVLP4F90/060 and recall — FDA Z-1643-2025

Device may have a small breach in the proximal end of the shaft under the strain relief of the delivery system. Breach is not visibly detectable and may…

Class IIclassification
Z-1643-2025official record ID
Apr 30, 2025report date

Recall record at a glance

Official recordZ-1643-2025
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmAbbott
Report dateApr 30, 2025
Recall initiation dateMar 17, 2025
Quantity4,850
LocationPlymouth, MN, United States

Official product description

Amplatzer TorqVue LP Delivery System (TVLP), Model nos. 9-TVLP4F90/060 and 9-TVLP4F90/080

Why the product was recalled

Device may have a small breach in the proximal end of the shaft under the strain relief of the delivery system. Breach is not visibly detectable and may lead to prolonged procedure, blood loss, or air ingress with potential for air embolism.

Distribution information

Worldwide - US Nationwide distribution.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.