FDA recall record

DICOM GRID Ambra PACS UDI: +AMBRHEALTHSOLUTIONS0/$$+7_V3.16.13.0R Software Version Number recall — FDA Z-1045-2017

A software error caused the window/level to become the same in one series regardless if the image had different levels; image results have a washed-out…

Class IIclassification
Z-1045-2017official record ID
Jan 25, 2017report date

Recall record at a glance

Official recordZ-1045-2017
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmDICOM GRID Inc.
Report dateJan 25, 2017
Recall initiation dateDec 16, 2016
Quantity209
LocationPhoenix, AZ, United States

Official product description

Ambra PACS UDI: +AMBRHEALTHSOLUTIONS0/$$+7_V3.16.13.0R Software Version Number: V3.16.13.0 Intended for use as a primary diagnostic and analysis tool for diagnostic images.

Why the product was recalled

A software error caused the window/level to become the same in one series regardless if the image had different levels; image results have a washed-out grey appearance.

Distribution information

U.S. distribution to the following; TX, FL, NE, CA, WI, AZ, NH, MO, MA, CO, PA, NV, PR, IL, OH, NY, MD, NJ, MN, MI, CT, VA, GA, UT, KS, WA, MI, HI. Foreign distribution to the following; Canada, Spain, Israel, Singapore, Kosovo, Austria, Brazil, Malaysia, Australia.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.