FDA recall record

DePuy Orthopaedics Altrx" Altralinked" Polyethylene liners Description: An acetabular liner recall — FDA Z-2220-2012

In the recalled lot, the products were not machined according to the production router, a size 52 cup (122136052) was machined instead. This resulted in a…

Class IIclassification
Z-2220-2012official record ID
Aug 29, 2012report date

Recall record at a glance

Official recordZ-2220-2012
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmDePuy Orthopaedics, Inc.
Report dateAug 29, 2012
Recall initiation dateJul 24, 2012
Quantity20 liners
LocationWarsaw, IN, United States

Official product description

Altrx" Altralinked" Polyethylene liners Description: An acetabular liner is one of the implants used during total hip arthroplasty to replace the damaged hip joint to help increase patient mobility and to reduce pain. Packaging: Product is placed into a PETG tray, sealed with a Tyvek lid. This is placed into a second PETG tray and sealed with a Tyvek lid. This package is placed into a paperboard carton, labeled, and shrink-wrapped. Intended Use: An acetabular liner is used in combination with other implants for total hip arthroplasty is intended to provide increased patient mobility and reduce pain by replacing the damaged hip joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components.

Why the product was recalled

In the recalled lot, the products were not machined according to the production router, a size 52 cup (122136052) was machined instead. This resulted in a size 52 liner being etched and labeled as a size 54 liner.

Distribution information

Worldwide Distribution - USA including PA and the country of Canada

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.