Recall record at a glance
| Official record | Z-0482-2025 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class I |
| FDA record status | Ongoing |
| Recalling firm | ICU Medical, Inc. |
| Report date | Nov 27, 2024 |
| Recall initiation date | Oct 22, 2024 |
| Quantity | 3840 |
| Location | Lake Forest, IL, United States |
Official product description
Allegedly counterfeit batteries manufactured by CSB batteries being used with Plum Infusion Systems: Plum 360" Infusion System, Pump List Number 30010; Plum A+ & Plum A+3" Infusion Systems, Pump List Numbers: 11005, 11971, 12391, 12618, 20678, 20679, 20792, 60529, 12348, 1197
Why the product was recalled
ICU Medical has received reports of allegedly counterfeit CSB batteries being used with Plum Infusion Systems. While these non-OEM batteries are visually similar to the Plum batteries supplied by ICU Medical, they are in fact not the same batteries and have not been tested or validated for use with Plum Infusion Systems. Preliminary reports suggest that these non-OEM batteries fail to hold their charge and the pump may display messages to replace batteries earlier than expected. OSI Batteries and their customers are distributing these non-OEM CSB batteries without authorization from ICU Medical.
Distribution information
Worldwide distribution: US (nationwide) including Puerto Rico and OUS (foreign) countries of: Argentina, Brazil, Canada, Chile, Colombia, France, Greece, Hong Kong, Italy, Ireland Jordan, Kuwait, Mexico, Netherlands, Oman, Peru, Philippines, Poland, Saudia Arabia, Spain, Taiwan, Turkey, United Kingdom, and Uruguay.
How to read the classification
FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.