FDA recall record

Abbott Molecular Alinity m System, REF 08N53-002, For In Vitro Diagnostic Use recall — FDA Z-0268-2023

Abbott has identified potential performance issues for the Alinity m System software version 1.6.5: 1) In a unique scenario, the waste chute flapper was…

Class IIclassification
Z-0268-2023official record ID
Nov 30, 2022report date

Recall record at a glance

Official recordZ-0268-2023
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmAbbott Molecular, Inc.
Report dateNov 30, 2022
Recall initiation dateOct 12, 2022
Quantity883 units
LocationDes Plaines, IL, United States

Official product description

Alinity m System, REF 08N53-002, For In Vitro Diagnostic Use

Why the product was recalled

Abbott has identified potential performance issues for the Alinity m System software version 1.6.5: 1) In a unique scenario, the waste chute flapper was found to not open when the Systems Solution drawer was closed and locked. It was found that the software will cause the flapper door to stay in the closed position when the drawer is closed and locked if the flapper had previously been moved out of position. 2) Sample preparation drawer #1 barcode information is used instead of sample prep drawer #2 when the scanned data is not sent to the System Control Center before the next bottle barcode in sample prep drawer#2 is scanned. 3) Under a specific condition, while the system is processing tests and a new test request is made, when the level of the bulk solution is too low to process a test, the software would error stop the system and try to complete all in-process tests. An internal counter is reset which can cause reaction vessels (RVs) already present in an Amplification Detection Unit (or amp detect) to remain there and not be moved to waste so that more RVs can be placed into the amp detect for processing. 4) 4 to 6 replicates of the same auto calibrator orders can be run with 2 different sets of calibrator materials on the same rack. It was discovered during internal testing, when running 4 to 6 replicates of a calibrator, it is possible for the user to use 2 separate lots of material which is not detected by the System Control Center (SCC). If this were to occur, the calibration curve would be created using 2 separate material lots. In normal use scenario, this curve would be made using only 1 lot of material.

Distribution information

Worldwide distribution.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.