FDA recall record

Abbott Laboratories Alinity ci-series System Control Module clinical chemistry and recall — FDA Z-2472-2021

Due to potential performance issues with software version 3.2.3 and earlier. Performance issues are: 1) There is the potential for onboard reagents past…

Class IIclassification
Z-2472-2021official record ID
Sep 22, 2021report date

Recall record at a glance

Official recordZ-2472-2021
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmAbbott Laboratories
Report dateSep 22, 2021
Recall initiation dateJul 26, 2021
Quantity4,833 systems
LocationIrving, TX, United States

Official product description

Alinity ci-series System Control Module clinical chemistry and immunoassay analyzer for in vitro diagnostic use. Model number LN 03R70-01

Why the product was recalled

Due to potential performance issues with software version 3.2.3 and earlier. Performance issues are: 1) There is the potential for onboard reagents past their lot expiration date or onboard stability time to be incorrectly displayed on the Reagent Status screen with a status of "OK". In this scenario, the expired reagent will remain in the carousel for sample processing. 2) Performing monthly maintenance procedure 5701 Clean ICT Drain Tip incorrectly may potentially cause issues such as damaged connector or leaking connection. 3) A sample exception with message code 150 "Unable to process test. Previous processing module error" is produced without notifying the operator of a reagent pipettor failure and may cause processing module to transition into the Pausing state. Tests initiated prior to the sample exception may continue to process without dispense of reagents, potentially leading to incorrect results. 4. Reagent short dispense: Alinity c needs a control measure to protect against the potential for short reagent dispense. Addition of a maximum depth limit for reagent cartridges on the Alinity c to prevent the potential for a short dispense is needed. 5. RSM indicators were blinking green incorrectly: The Reagent and Sample Manager (RSM) indicators will incorrectly blink green indicating that processing is completed and the rack or cartridge can be removed from the loading area when the rack or cartridge should not be accessed. 6. Erroneous ICT Calibration Curves: Erroneous ICT calibration curves might be generated for the Alinity c when there is a malfunction of ICT reference solution delivery, use of wrong/evaporated calibrator, defective ICT modules, etc.

Distribution information

Worldwide distribution. US nationwide including Puerto Rico; Albania, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bermuda, Bosnia and Herzegovina, Botswana, Brazil, Brunei, Bulgaria, Burkina Faso, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Egypt, Estonia, Ethiopia, Finland, France, Germany, Greece, Guatemala, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Latvia, Lebanon, Lithuania, Madagascar, Malaysia, Maldives, Mauritius, Mexico, Moldova, Montenegro, Morocco, Namibia, Netherlands, New Caledonia, New Zealand, Nicaragua, Norway, Oman, Pakistan, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad & Tobago, Tunisia, Turkey, United Arab Emirates, United Kingdom, Ukraine, Uruguay, Vietnam and Zambia.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceNov 19, 2025

Alinity hq Analyzer recall — FDA Z-0489-2026

Software issue for hq analyzer results in system not visibly applying appropriate flagging to results when saturation is present which may result in incorrect results.

Class IIFDA: OngoingAbbott Laboratories
Medical DeviceApr 30, 2025

CELL-DYN Ruby recall — FDA Z-1652-2025

When expired reagents are scanned or manually entered, the system will change the expiration date to current or future date without notifying user that an expired reagent is being used. Use…

Class IIFDA: OngoingAbbott Laboratories
Medical DeviceApr 30, 2025

CELL-DYN Ruby recall — FDA Z-1654-2025

When expired reagents are scanned or manually entered, the system will change the expiration date to current or future date without notifying user that an expired reagent is being used. Use…

Class IIFDA: OngoingAbbott Laboratories
Medical DeviceApr 30, 2025

CELL-DYN Ruby recall — FDA Z-1655-2025

When expired reagents are scanned or manually entered, the system will change the expiration date to current or future date without notifying user that an expired reagent is being used. Use…

Class IIFDA: OngoingAbbott Laboratories
Medical DeviceApr 30, 2025

CELL-DYN Ruby recall — FDA Z-1656-2025

When expired reagents are scanned or manually entered, the system will change the expiration date to current or future date without notifying user that an expired reagent is being used. Use…

Class IIFDA: OngoingAbbott Laboratories