FDA recall record

Abbott Gmbh & Co. KG Alinity c Cuvette Segments, List Number 04S47-01 Product Usage: The recall — FDA Z-1709-2019

Individual cuvettes within the Alinity c Cuvette Segment may become seated lower than the designed height. This may result in inadequate dispense into…

Class IIclassification
Z-1709-2019official record ID
Jun 12, 2019report date

Recall record at a glance

Official recordZ-1709-2019
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmAbbott Gmbh & Co. KG
Report dateJun 12, 2019
Recall initiation dateApr 15, 2019
Quantity19,184
LocationWiesbaden, Germany

Official product description

Alinity c Cuvette Segments, List Number 04S47-01 Product Usage: The Alinity c Cuvettes are used on the Alinity c system processing module. The Alinity c processing module is a fully automated chemistry analyzer allowing random and continuous access, as well as priority and automated retest processing using photometric and potentiometric detection technology.

Why the product was recalled

Individual cuvettes within the Alinity c Cuvette Segment may become seated lower than the designed height. This may result in inadequate dispense into specific cuvettes due to the sample probe being unable to make efficient contact with the cuvette bottom.

Distribution information

US Nationwide Distribution- AL, AR,CA, FL, GA, IL, LA, MA, ME, MN, MO, NJ, NY, OK, SC, TN, TX, UT, & VA WORLDWIDE - Albania, Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, Columbia, Croatia, Czech Republic, Estonia, Finland, France, Germany, Greece, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Jordan, Kenya, Kuwait, Latvia, Lebanon, Lithuania, Malaysia, Mexico, Netherlands, Norway, Pakistan, Philippines, Poland, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, Spain, Sweden, Thailand, Trinidad & Tobago, Turkey, United Kingdom, and Vietnam.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.