FDA recall record

Abbott GmBH & Co. KG Alinity c Carbon Dioxide Reagent Kit, Produced for Abbott by Sekisui recall — FDA Z-0575-2020

Abbott internal studies have determined that atmospheric Carbon Dioxide (CO2) can be absorbed into the Alinity c Carbon Dioxide reagent, resulting in the…

Class IIclassification
Z-0575-2020official record ID
Dec 4, 2019report date

Recall record at a glance

Official recordZ-0575-2020
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmAbbott GmBH & Co. KG
Report dateDec 4, 2019
Recall initiation dateOct 23, 2019
Quantity105 kits (lot 54449UQ03) and 1215 kits (lot 54454UQ03)
LocationWiesbaden, Germany

Official product description

Alinity c Carbon Dioxide Reagent Kit, Produced for Abbott by Sekisui Diagnostics P.E.I. Inc. Prince Edward Island Canada

Why the product was recalled

Abbott internal studies have determined that atmospheric Carbon Dioxide (CO2) can be absorbed into the Alinity c Carbon Dioxide reagent, resulting in the potential for incorrect results. Abbott was made aware of this issue from customer complaints. Internal testing demonstrated that the amount of CO2 absorbed is higher with increased reagent carousel rotation and when the volume of reagent in the cartridge is reduced. This phenomenon can be detected as a shift in QC. The shift is predictive in both direction and magnitude of the potential impact to patient results.

Distribution information

Worldwide and US Nationwide in the states of: AL, AR, CA, CO, FL, GA, ID, IL, LA, MA, MI, MN, MO, NC, NY, OK, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, and Puerto Rico Foreign: AUSTRALIA, BANGLADESH, BELGIUM, BOTSWANA, BRAZIL, CANADA, COLOMBIA, DENMARK, FRANCE, GREECE, Hong Kong, INDIA, ITALY, JAMAICA, LEBANON, MAURITIUS, NAMIBIA, NETHERLANDS, NEW ZEALAND, PAKISTAN, PHILIPPINES, POLAND, PORTUGAL, PORTUGAL, ROMANIA, SAUDI ARABIA, SINGAPORE, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, THAILAND, TRINIDAD & TOBAGO, UAE, and UNITED KINGDOM

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.