FDA recall record

Alere Triage D-Dimer Test PN 98100, Lot W53884B recall — FDA Z-1727-2014

Alere initiated this recall because a limited number of Alere Triage D-Dimer devices from PN 98100 Lot W53884B were incorrectly manufactured. As a result,…

Class IIclassification
Z-1727-2014official record ID
Jun 18, 2014report date

Recall record at a glance

Official recordZ-1727-2014
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmAlere San Diego, Inc.
Report dateJun 18, 2014
Recall initiation dateMay 2, 2014
Quantity373 kits
LocationSan Diego, CA, United States

Official product description

Alere Triage D-Dimer Test PN 98100, Lot W53884B. The Alere Triage D-Dimer Test is a fluorescence immunoassay to be used with the Alere Triage Meters for the quantitative determination of cross-linked fibrin degradation products containing D-dimer in EDTA anticoagulated whole blood and plasma specimens. . The Alere Triage D-Dimer test is intended for use as an aid in the assessment and evaluation of patients suspected of having disseminated intravascular coagulation or thromboembolic events including pulmonary embolism.

Why the product was recalled

Alere initiated this recall because a limited number of Alere Triage D-Dimer devices from PN 98100 Lot W53884B were incorrectly manufactured. As a result, when testing patient sample these affected devices will either generate an error code (e.g. E4 error), or they may generate an incorrect result.

Distribution information

Nationwide Distribution only.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.