FDA recall record

Alcon Monarch III D Delivery Cartridge recall — FDA Z-0835-2025

Increase in complaints associated with a molding defect within the inner lumen of delivery cartridges which could result in a scratch on the intraocular…

Class IIclassification
Z-0835-2025official record ID
Jan 22, 2025report date

Recall record at a glance

Official recordZ-0835-2025
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmAlcon Research LLC
Report dateJan 22, 2025
Recall initiation dateNov 13, 2024
Quantity41718 units
LocationFort Worth, TX, United States

Official product description

Brand Name: Alcon Product Name: Monarch III D Delivery Cartridge Model/Catalog Number: 8065977763 Product Description: Monarch III D delivery cartridge Lens Guide, Intraocular Component: N/A

Why the product was recalled

Increase in complaints associated with a molding defect within the inner lumen of delivery cartridges which could result in a scratch on the intraocular lens (IOL) upon delivery.

Distribution information

US: AK AL AR AZ CA CO CT DE FL GA GU HI IA ID IL IN KS KY LA MA MD ME MI MN MO MS MT NC ND NE NH NJ NM NV NY OH OK OR PA RI SC SD TN TX UT VA VT WA WI WV WY OUS: Russia, Egypt

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

DrugMay 6, 2026

SYSTANE recall — FDA D-0491-2026

Lack of Assurance of Sterility: Due to FDA inspection observations that it believes may impact product quality.

Class IIFDA: OngoingAlcon Research LLC