FDA recall record

CareFusion 303 Alaris PC unit model 8015, Infusion pump. software version 9.17 Product recall — FDA Z-1811-2015

An issue with the cancel functionality used during atypical infusion programming to cancel user inputted values. An infusion may start that is greater…

Class IIclassification
Z-1811-2015official record ID
Jun 24, 2015report date

Recall record at a glance

Official recordZ-1811-2015
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmCareFusion 303, Inc.
Report dateJun 24, 2015
Recall initiation dateMay 13, 2015
Quantity63 units
LocationSan Diego, CA, United States

Official product description

Alaris PC unit model 8015, Infusion pump. software version 9.17 Product Usage: The Alaris PC unit is the central programming, monitoring and power supply component for the Alaris System. The software is embedded in the Alaris PC unit.

Why the product was recalled

An issue with the cancel functionality used during atypical infusion programming to cancel user inputted values. An infusion may start that is greater than or less than the hospital established limits for the specific medication.

Distribution information

US Nationwide Distribution

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records