FDA recall record

CareFusion 303 Alaris PC Unit, Infusion Pump Model 8000, Part No recall — FDA Z-1239-2016

CareFusion is recalling the Alaris PC unit because a component on the PC unit power supply may cause a "System Error" or "Missing Battery" error code…

Class IIclassification
Z-1239-2016official record ID
Mar 30, 2016report date

Recall record at a glance

Official recordZ-1239-2016
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmCareFusion 303, Inc.
Report dateMar 30, 2016
Recall initiation dateFeb 12, 2016
Quantity170 units
LocationSan Diego, CA, United States

Official product description

Alaris PC Unit, Infusion Pump Model 8000, Part No. TC10005092

Why the product was recalled

CareFusion is recalling the Alaris PC unit because a component on the PC unit power supply may cause a "System Error" or "Missing Battery" error code (120.4630).

Distribution information

Nationwide in US: AR, AZ, CA, FL, GA, IN, KY, MD, MI, MN, MO, NJ, NM, NY, OH, OR, PA, SC, TN, TX, VA, WI.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.