FDA recall record

BMES Alaris Infusion Pump Module, Model 8100, serviced/repaired by Biomedical Equipment recall — FDA Z-2204-2021

Bezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of…

Class Iclassification
Z-2204-2021official record ID
Aug 18, 2021report date

Recall record at a glance

Official recordZ-2204-2021
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusTerminated
Recalling firmBioMedical Equipment Service Co (BMES)
Report dateAug 18, 2021
Recall initiation dateFeb 25, 2021
Quantity718 potentially affected pumps
LocationLouisville, KY, United States

Official product description

Alaris Infusion Pump Module, Model 8100, serviced/repaired by Biomedical Equipment Service Co. (BMES) with an impacted bezel repair part TIPA-8100-4410

Why the product was recalled

Bezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel repair part posts may result in free flow, over infusion, under infusion or interruption of infusion.

Distribution information

The affected products were distributed to the following US states: CA, IL, IN, KY, LA, NJ, NV, OH, and PA.

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.