FDA recall record

Vyaire Medical AirLife Adult Manual Resuscitator, 40" (1.0 m) Oxygen Reservoir Tubing, Adult recall — FDA Z-1063-2024

Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device…

Class Iclassification
Z-1063-2024official record ID
Mar 6, 2024report date

Recall record at a glance

Official recordZ-1063-2024
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusOngoing
Recalling firmVyaire Medical
Report dateMar 6, 2024
Recall initiation dateJan 10, 2024
Quantity4,572 units
LocationMettawa, IL, United States

Official product description

AirLife Adult Manual Resuscitator, 40" (1.0 m) Oxygen Reservoir Tubing, Adult Mask, CO2 Detector, REF 2K8005C2

Why the product was recalled

Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.

Distribution information

Worldwide distribution. US nationwide, Brazil, Canada, Switzerland, France, UK, Mexico, Netherlands, Saudi Arabia, Sweden, and South Africa

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceMay 21, 2025

AirLife Infant Heated Wire Circuit recall — FDA Z-1697-2025

Adapters may disconnect during setup or while in use, potentially interrupting ventilation. Any disconnection can result in severe consequences for neonates, including hypoxia, hypercapnia,…

Class IFDA: OngoingVyaire Medical
Medical DeviceMay 21, 2025

AirLife Infant Heated Wire Circuit recall — FDA Z-1698-2025

Adapters may disconnect during setup or while in use, potentially interrupting ventilation. Any disconnection can result in severe consequences for neonates, including hypoxia, hypercapnia,…

Class IFDA: OngoingVyaire Medical