AGFA Digital Radiography X-Ray system DX-D 100 Agfa's DX-D 100 is indicated for use in recall — FDA Z-2081-2014

When liquid comes in contact with the DX-D 100 touch screen, the device may incorrectly recognize this as user input altering device settings.

Class IIclassification
Z-2081-2014official record ID
Jul 30, 2014report date

Recall record at a glance

Official recordZ-2081-2014
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmAGFA Corp.
Report dateJul 30, 2014
Recall initiation dateJun 6, 2014
Quantity216 units
LocationGreenville, SC, United States

Official product description

AGFA Digital Radiography X-Ray system DX-D 100 Agfa's DX-D 100 is indicated for use in providing diagnostic quality images to aid the physician with diagnosis. Systems can be used with MUSICA2 image processing to create radiographic images of the skeleton including skull, spinal column and extremities: chest, abdomen, and other body parts. Agfa's DX-D 100 is not indicated for use in mammography.

Why the product was recalled

When liquid comes in contact with the DX-D 100 touch screen, the device may incorrectly recognize this as user input altering device settings.

Distribution information

USA (nationwide) including the states of AL, CA, FL, GA, IL, IN, IA, KY, LA, MA, MI, MN, MS, MO, NE, NJ, NY, NC, OH, SC, TN, TX, VA, WV, WI, WY and Internationally to Canada.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.