FDA recall record

Hand Biomechanics Lab Agee-WristJack Fracture Reduction System recall — FDA Z-0136-2017

A number of Agee WristJack Fracture Reduction Systems have been manufactured with a molded beam slightly larger than optimal. This may result in the…

Class IIclassification
Z-0136-2017official record ID
Oct 26, 2016report date

Recall record at a glance

Official recordZ-0136-2017
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmHand Biomechanics Lab Inc.
Report dateOct 26, 2016
Recall initiation dateSep 14, 2016
Quantity65 units
LocationSacramento, CA, United States

Official product description

Agee-WristJack Fracture Reduction System; Item number: CFD-147. The WristJack Fracture Reduction System is a fracture reduction and external fixation system for treatment of distal radius fractures. This is a single use system used for a trauma related surgery The WristJack Fracture Reduction System is a fracture reduction and external fixation system for treatment of distal radius fractures. The non-perishable sterile system is packaged in an inner then outer tray and shipped in a cardboard shelf box. Orthopedic:The WristJack Fracture Reduction System is a fracture reduction and external fixation system for treatment of distal radius fractures.

Why the product was recalled

A number of Agee WristJack Fracture Reduction Systems have been manufactured with a molded beam slightly larger than optimal. This may result in the surgeon experiencing difficulty fine-tuning the length (red) and radial/ulnar alignment (blue) adjustments on the device.

Distribution information

Nationwide Distribution

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.