FDA recall record

AESDEX recall — FDA Z-2433-2018

Following pressurization of the devices, damage to the device handle mechanism may occur, accompanied by a loud noise and potential dispersion of plastic…

Class IIclassification
Z-2433-2018official record ID
Jul 25, 2018report date

Recall record at a glance

Official recordZ-2433-2018
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmAESDEX
Report dateJul 25, 2018
Recall initiation dateApr 2, 2018
Quantity587 units
LocationRedwood City, CA, United States

Official product description

AesDex, LLC Cardica C-Port xA PLUS Anastomosis System. Catalog Number FG-000100. The Cardica C-Port xA PLUS Anastomosis System, delivers a series of clips that create an anastomosis between a small target vessel (e.g. coronary artery) and conduit (e.g. saphenous vein graft). The stainless steel clips create a complete end-to-side anastomosis that is functionally equivalent to a hand-sutured interrupted stitch anastomosis. The system consists of one Anastomosis Device and one Retractor Mount.

Why the product was recalled

Following pressurization of the devices, damage to the device handle mechanism may occur, accompanied by a loud noise and potential dispersion of plastic fragments and components.

Distribution information

U.S. and international: Belgium, Germany, Italy and Netherlands. No gov accounts.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceJul 25, 2018

AESDEX recall — FDA Z-2434-2018

Following pressurization of the devices, damage to the device handle mechanism may occur, accompanied by a loud noise and potential dispersion of plastic fragments and components.

Class IIFDA: TerminatedAESDEX