FDA recall record

Aesculap GNI161 Bipolar Foot Control Product Usage: The Aesculap Bipolar Coagulator is recall — FDA Z-0479-2014

Complaints were received for the GN161 Bipolar Foot Control reporting that the bipolar energy did not stop after release of the foot control.

Class IIclassification
Z-0479-2014official record ID
Dec 18, 2013report date

Recall record at a glance

Official recordZ-0479-2014
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmAesculap, Inc.
Report dateDec 18, 2013
Recall initiation dateSep 10, 2013
Quantity85
LocationCenter Valley, PA, United States

Official product description

Aesculap GNI161 Bipolar Foot Control Product Usage: The Aesculap Bipolar Coagulator is intended for use in surgery to generate electrical power for bipolar instruments. Bipolar coagulators are used in neurosurgery, ENT surgery, urology, laparoscopy, and plastic surgery.

Why the product was recalled

Complaints were received for the GN161 Bipolar Foot Control reporting that the bipolar energy did not stop after release of the foot control.

Distribution information

Worldwide Distribution - USA Nationwide and countries of: Austria, Germany, Japan, and Taiwan.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.