Recall record at a glance
| Official record | Z-0815-2015 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Philips Ultrasound, Inc. |
| Report date | Dec 24, 2014 |
| Recall initiation date | Nov 5, 2014 |
| Quantity | additional 52 international units; 261 units (73 in the US and 188 outside the US); 417 units (289 in the US and 128 outside US) |
| Location | Bothell, WA, United States |
Official product description
Additional devices identified in new FSN MA-FCO 83000190: QLAB versions 10.0 with a2DQ and/or aCMQ plug-ins installed with Philips Xcelera software. *********************************************************************************** Devices identified in the Field Safety Notice ( FSN MA-FCO79500316 ) QLAB Quantification Software. QLAB version 10.0, 10.1 and 10.1.1 with a2DQ and/or aCMQ plug-ins are affected. Catalog # 795041 and Part # 989605315491 Designed to view and quantify image data acquired on Philips Healthcare ultrasound products.
Why the product was recalled
When using the QLAB Auto 2D Quantification (a2DQ) and Auto Cardiac Motion Quantification (aCMQ) applications, the reported End-Systolic Volume (ESV) may be smaller, and the Left Ventricular Ejection Fraction (EF) may be higher, than the ESV and EF calculated by manual tracing without the use of QLAB. An incorrect EF calculation could lead to misdiagnosis and/or delayed or incorrect therapy.
Distribution information
Worldwide distribution (for FCO 83000190): US, including the states of AK, CA, CO, CT, DC, DE, FL, GA, ID, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, NC, NJ, NY, OH, OR, PA, SC, SD, TN, TX, VA, WA, and WI; and the countries of United Arab Emirates, Austria, Australia, Belgium, Canada, Switzerland, Colombia, Germany, Ecuador, Estonia, Spain, Finland, France, Great Britain, Hong Kong, Israel, Italy, Japan, Luxembourg, Malaysia, Netherlands, Norway, New Zealand, Philippines, Saudi Arabia, Sweden, Singapore, Slovenia, Thailand, Turkey, Taiwan, and Vietnam. ********************************************************************************** Worldwide Distribution -- US, including the states of KY, TN, ME, CA, PA, NC, CT, NY, SC, IL, IN, and WI; and the countries of Argentina, Spain, India, Italy, Austria, Germany, Belgium, Canada, Finland, Switzerland, Greece, Malaysia, Sweden, Australia, Netherlands, Mexico, Peru, Taiwan, Japan, France, and Norway.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.