FDA recall record

Ion Beam Applications S.A adaPT insight (12C), v1.3.2, PAT.109 (US), Treatment room 4: I2C recall — FDA Z-0836-2015

Two issues were identified: 1. In the release of adaPT insight v1.3.2., the kV/kV Single Source acquisition workflow was linked to a wrong geometrical…

Class IIclassification
Z-0836-2015official record ID
Dec 31, 2014report date

Recall record at a glance

Official recordZ-0836-2015
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmIon Beam Applications S.A.
Report dateDec 31, 2014
Recall initiation dateNov 7, 2014
Quantity1
LocationLouvain La Neuve, Belgium

Official product description

adaPT insight (12C), v1.3.2, PAT.109 (US), Treatment room 4: I2C is used with a charged particle or photon radiation therapy system for localization of the patient position with respect to the therapy equipment and to provide correction feedback to the radiation therapy device.

Why the product was recalled

Two issues were identified: 1. In the release of adaPT insight v1.3.2., the kV/kV Single Source acquisition workflow was linked to a wrong geometrical calibration file, resulting in an alignment offset for this specific mode of acquisition. 2. A gantry collision incident required the IBA service team to verify the gantry alignment and protons I x-ray beams colinearity, du

Distribution information

US Distribution in PA only.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.