Recall record at a glance
| Official record | Z-2396-2019 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Ad-Tech Medical Instrument Corporation |
| Report date | Sep 4, 2019 |
| Recall initiation date | Apr 16, 2019 |
| Quantity | 18 |
| Location | Oak Creek, WI, United States |
Official product description
AD-TECH Spencer Probe Depth Electrode Product Usage: The Ad-Tech Depth Electrodes (Depth Electrodes, Foramen Ovale Depth Electrodes, Macro Micro Depth Electrodes, Spencer Probe Depth Electrodes, Wyler Sphenoidal Depth Electrodes) are intended for temporary (< 30 days) use with recording, monitoring and stimulation equipment for the recording, monitoring and stimulation of electrical signals at the subsurface level of the brain. The recording of electrical activity supports definition of the location of epileptogenic foci and brain mapping.
Why the product was recalled
An Ad-Tech Clinical Specialist, attended a case on April 15, 2019. During the case the surgeon encountered an issue after implanting four Ad-Tech depth electrodes. Upon initial device testing subsequent to implantation it was identified that two of the devices that were labeled as 8 contact depth electrodes (SD08R-SP05X-000) were in fact 6 contact depth electrodes (SD06R-SP05X-000) . There was no impact to the patient due to this incident.
Distribution information
Worldwide - Nationwide distribution in the state of Illinois and countries of Russia and United Kingdom
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.