Recall record at a glance
| Official record | Z-2352-2019 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Ongoing |
| Recalling firm | Siemens Medical Solutions USA, Inc. |
| Report date | Aug 28, 2019 |
| Recall initiation date | Jan 28, 2019 |
| Quantity | 1384 systems |
| Location | Mountain View, CA, United States |
Official product description
ACUSON NX3 Diagnostic Ultrasound System, Model # 11235612, RX Only, Siemens Medical Solutions, containing software versions: VA10A, VA10B, VA10C, VA10D, VA10E, VA10F, VB20A, and VB20B; UDI # 04056869004297 The ACUSON NX2 diagnostic ultrasound imaging systems are intended for the following applications: Cardiac, Fetal, Abdominal (including liver), Pediatric, Small Parts (Small Organ), Transcranial, OB/GYN, Pelvic, Urology, Vascular (including Peripheral Vessel) and Musculoskeletal applications. The systems also provide for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes.
Why the product was recalled
The action is being initiated due to internal testing which identified a possibility for transducers to exceed the acoustic output power (AOP) values defined . The transit voltage values based on PW Doppler are calculated normally but not loaded correctly into hardware
Distribution information
Worldwide Distribution: US (nationwide) including states of: AL, AZ, AR, CA, CO, CT, DC, FL, GA, IL,IA, IN, KY, LA, MD, MA, MI, MN, MS, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI and WY; and countries of: Albania, Algeria, Angola, Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia, Brazil, Brunei, Bulgaria, Burkina Faso, Cameroon, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyrus, Czech Republic, Democratic Republic of Congo, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador Finland, France, Gabon, Georgia, Germany, Greece, Guatemala, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Ivory Coast, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Latvia, Lebanon, Luxembourg, Macedonia, Malaysia, Maldives, Mali, Mexico, Morocco, Nepal, Netherlands, Norway, Pakistan, Palestine, Panama, Philippines, Poland, Portugal, Republic of Korea, Romania, Ruanda, Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Turkey, U.A.E., Uganda, Ukraine, United Kingdom, Uruguay, Uzbekistan and Vietnam.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.