Recall record at a glance
| Official record | Z-2236-2024 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Ongoing |
| Recalling firm | Exactech, Inc. |
| Report date | Jul 10, 2024 |
| Recall initiation date | Apr 26, 2024 |
| Quantity | 589 units |
| Location | Gainesville, FL, United States |
Official product description
Acumatch L-Series BIPOLAR LINER, 22mm (BIPOLAR ENDOPROSTHESIS-Hip), Item Numbers: a) 100-22-19, SZ. L; b) 100-22-20, SZ. M; c) 100-22-21, SZ. N; d) 100-22-22, SZ. P; e) 100-22-23, SZ. R; f) 100-22-24, SZ. S; g) 100-22-25, SZ. T
Why the product was recalled
The AcuMatch L-Series 22mm Inner Diameter Bipolar Hip Liner lots were packaged without the specified ethylene vinyl alcohol (EVOH) layer. Between 2004 and August 2021, our packaging process utilized two different types of packaging materials: 1) Low Density Polyethylene (LDPE), Nylon, and EVOH, or 2) LDPE and Nylon without EVOH.
Distribution information
Worldwide - US distribution in Puerto Rico and the countries of Great Britain, Japan.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.