FDA recall record

GE Healthcare Achilles Express Bone Sonometer recall — FDA Z-0113-2020

Certain Achilles Express and Achilles Insight systems were shipped with European CEE-7/7 Type Power Cords to countries where three protruding pin plugs…

Class IIclassification
Z-0113-2020official record ID
Oct 16, 2019report date

Recall record at a glance

Official recordZ-0113-2020
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmGE Healthcare, LLC
Report dateOct 16, 2019
Recall initiation dateSep 3, 2019
Quantity241 devices
LocationWaukesha, WI, United States

Official product description

Achilles Express Bone Sonometer

Why the product was recalled

Certain Achilles Express and Achilles Insight systems were shipped with European CEE-7/7 Type Power Cords to countries where three protruding pin plugs are required for protective earth connection.

Distribution information

Nationwide distribution to AZ, GA, MT, NC, NJ, NY, OH, WI. International distribution to Australia, Bahrain, Bangladesh, Brazil, Cambodia, Chile, China, Colombia, Cyprus, Hong Kong, India, Iraq, Ireland, Italy, Kuwait, Lebanon, Malaysia, Mauritius, Mexico, MEXICO, Myanmar, Nepal, Oman, Panama, Peru, Philippines, Saudi Arabia, Singapore, Sri Lanka, Switzerland, Taiwan, Thailand, United Arab Emirates, United Kingdom, Uruguay, Vietnam

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.