Zimmer Biomet ACE Trochanteric Nail System STER TROC END CAP STER TROC END CAP 5MM STER TROC recall — FDA Z-1226-2018

Zimmer Biomet is conducting a medical device recall for the ACE Trochanteric Nail (ATN) system after a Device Performance Review (DPR) failed clinical…

Class IIclassification
Z-1226-2018official record ID
Apr 4, 2018report date

Recall record at a glance

Official recordZ-1226-2018
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmZimmer Biomet, Inc.
Report dateApr 4, 2018
Recall initiation dateDec 20, 2017
Quantity3454
LocationWarsaw, IN, United States

Official product description

ACE Trochanteric Nail System STER TROC END CAP STER TROC END CAP 5MM STER TROC END CAP 10MM STER TROC END CAP 15MM STER TROC END CAP 20MM STER TROC END CAP 25MM The Trochanteric Nail System is intended to treat stable and unstable proximal fractures of the femur including pertrochanteric fractures, intertrochanteric fractures, high subtrochanteric fractures and combinations of these fractures. The Trochanteric Long Nail System is additionally indicated to treat pertrochanteric fractures associated with shaft fractures, pathologic fractures in osteoporotic bone of the trochanteric and diaphyseal areas, long subtrochanteric fractures, ipsilateral femoral fractures, proximal or distal nonunions and malunions and revision procedures.

Why the product was recalled

Zimmer Biomet is conducting a medical device recall for the ACE Trochanteric Nail (ATN) system after a Device Performance Review (DPR) failed clinical analysis for survivorship pursuant to the acceptance criteria for the DPR protocol for the ATN system.

Distribution information

USA (nationwide)

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.