FDA recall record

Angiodynamics Accu2i pMTA Applicator, 14 cm STANDARD, REF/Catalog # 900-600-US, Model # recall — FDA Z-0529-2018

The firm is recalling product that has the potential to not function during use. The product affected by this recall will not deliver the desired…

Class IIclassification
Z-0529-2018official record ID
Feb 14, 2018report date

Recall record at a glance

Official recordZ-0529-2018
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmAngiodynamics, Inc.
Report dateFeb 14, 2018
Recall initiation dateJun 5, 2017
Quantity14,353 units in total
LocationQueensbury, NY, United States

Official product description

Accu2i pMTA Applicator, 14 cm STANDARD, REF/Catalog # 900-600-US, Model # (UPN) H787900600US0, STERILE --- This is only for use with a Sulis V Generator, REF 806-10. Product Usage: The Acculis Accu2i pMTA Microwave Tissue Ablation Applicator is indicated for the intraoperative coagulation of soft tissue.

Why the product was recalled

The firm is recalling product that has the potential to not function during use. The product affected by this recall will not deliver the desired microwave energy due to coolant ingress in an electrical connection; this condition results in an error code High Reflective Power displayed on the microwave generator.

Distribution information

Worldwide distribution - US Nationwide

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.