FDA recall record

Roche Diabetes Care Accu-Chek Connect Diabetes Management App versions 2.0.0, 2.0.1 and recall — FDA Z-1899-2017

A program error (bug) in the Bolus Advisor feature, which could result in incorrect bolus advice and a potential insulin over-delivery.

Class IIclassification
Z-1899-2017official record ID
May 10, 2017report date

Recall record at a glance

Official recordZ-1899-2017
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmRoche Diabetes Care, Inc.
Report dateMay 10, 2017
Recall initiation dateMar 14, 2017
Quantity64,324
LocationIndianapolis, IN, United States

Official product description

Accu-Chek Connect Diabetes Management App versions 2.0.0, 2.0.1 and 2.1.0 for iOS and Android Product Usage: This is a digital product available for download direct to customers from the Apple App Store and the Google Play Store. The total number of downloads is not available. The number of unique users for each affected version from the date of initial distribution through 03/23/2017 is noted in 7a. However, some users will have used all three affected versions as they upgraded from version to version when the versions became available and many others will have tried the app and discontinued use. In the last 30 days, 02/22/2017 03/23/2017, there have been 16,781 unique users in the US with 264 of those seeking bolus advice and 19,268 unique users in the rest of world (excluding Canada) with 415 of those users seeking bolus advice.

Why the product was recalled

A program error (bug) in the Bolus Advisor feature, which could result in incorrect bolus advice and a potential insulin over-delivery.

Distribution information

US Nationwide in the states of: AL, AR, AZ, CA, CO, CT, DE, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Other.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.