Recall record at a glance
| Official record | Z-2185-2018 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Roche Diabetes Care, Inc. |
| Report date | Jun 20, 2018 |
| Recall initiation date | May 7, 2018 |
| Quantity | 295, 248 |
| Location | Indianapolis, IN, United States |
Official product description
Accu-Chek Aviva Plus Test Strip, Blood glucose monitoring system, Catalog nos. 06908217001, 06908268001, 06908349001,06908217001 Product Usage: The Accu-Chek Aviva Plus test strips are for use with the Accu-Chek Aviva Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper arm, or palm.
Why the product was recalled
The recalling firm identified four US lots of test strips out of specification. Affected strips show increased potential for strip errors prior to applying blood, strips not detected once inserted, or biased results, which may not be detectable. In the case of a biased result, incorrect action could be taken by the user in response to the displayed blood glucose result.
Distribution information
Worldwide Distribution - US Nationwide distribution. Foreign distribution to Argentina Armenia Australia Austria Azerbaijan Bahamas Bahrain Barbados Belgium Brazil Bulgaria Canada Chile China Colombia Costa Rica Denmark Dominican Republic El Salvador Finland Georgia Germany Ghana Greece Guatemala Honduras Hong Kong (SAR China) Hungary India Indonesia Israel Italy Jamaica Japan Jordan Ireland Kenya Korea Kuwait Lebanon Lithuania Luxembourg Malaysia Malta Mexico Myanmar Netherlands New Zealand Norway Oman Pakistan Philippines Poland Russian Federation New Zealand Saudi Arabia Serbia Singapore Slovenia South Africa Spain Sweden Switzerland Taiwan (Republic of China), Thailand, United Arab Emirates, United Kingdom, Vietnam, Portugal"
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.