FDA recall record

Beckman Coulter Access Thyroglobulin Antibody II, REF A32898, containing Instructions for recall — FDA Z-0265-2025

IFUs for the device incorrectly states the biotin interference was remediated. The issue could result in potentially false-low results.

Class IIclassification
Z-0265-2025official record ID
Nov 6, 2024report date

Recall record at a glance

Official recordZ-0265-2025
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmBeckman Coulter, Inc.
Report dateNov 6, 2024
Recall initiation dateSep 5, 2024
Quantity465,481 kits
LocationChaska, MN, United States

Official product description

Access Thyroglobulin Antibody II, REF A32898, containing Instructions for Use, IFU REF B14994, B63978, and A38168. The Access Thyroglobulin Antibody II (TgAb) assay is used for the quantitative determination of thyroglobulin antibody levels in human serum and plasma using the Access Immunoassay Systems.

Why the product was recalled

IFUs for the device incorrectly states the biotin interference was remediated. The issue could result in potentially false-low results.

Distribution information

Distribution was made nationwide to the U.S., including Puerto Rico. There was government distribution but no military distribution. Foreign distribution was made to Mexico.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.