FDA recall record

Beckman Coulter Access Intact PTH assay on the DxI 9000 Access Immunoassay Analyzer, Catalog recall — FDA Z-0673-2025

The reason for the recall is that EDTA and lithium heparin plasma samples that are stored in the DxI 9000 Access Immunoassay Analyzer sample wheel for a…

Class IIclassification
Z-0673-2025official record ID
Dec 18, 2024report date

Recall record at a glance

Official recordZ-0673-2025
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmBeckman Coulter Inc.
Report dateDec 18, 2024
Recall initiation dateNov 8, 2024
Quantity31 US; 167 OUS
LocationBrea, CA, United States

Official product description

Access Intact PTH assay on the DxI 9000 Access Immunoassay Analyzer, Catalog number A16972, for the quantitative determination of intact parathyroid hormone (parathyrin, PTH) levels in human serum and plasma.

Why the product was recalled

The reason for the recall is that EDTA and lithium heparin plasma samples that are stored in the DxI 9000 Access Immunoassay Analyzer sample wheel for a prolonged period before performing the intact PTH (parathyroid hormone) test may produce falsely decreased intact parathyroid hormone results. This could lead to a missed diagnosis, a delay in treatment, or inappropriate clinical management, depending upon the patient s clinical condition.

Distribution information

Foreign distribution to Albania, Australia, Austria, Bahrain, Belgium, Brazil, C¿te d'Ivoire, Croatia, Czechia, Egypt, France, Germany, Ghana, Hungary, India, Ireland, Israel, Italy, Kazakhstan, Kuwait, Malaysia, Netherlands, New Zealand, Poland, Portugal, Slovakia, South Africa, S. Korea, Spain, Switzerland, Taiwan, United Kingdom of Great Britain and Northern Ireland.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.