Beckman Coulter Access 2 Immunoassay System, Remanufactured, Catalog No. 386220 is an recall — FDA Z-1956-2016

Due to the lack of updated thermal components, these instruments may be more susceptible to the internal case temperature exceeding the 18 to 36 degrees…

Class IIclassification
Z-1956-2016official record ID
Jun 22, 2016report date

Recall record at a glance

Official recordZ-1956-2016
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmBeckman Coulter Inc.
Report dateJun 22, 2016
Recall initiation dateMay 17, 2016
Quantity4 units total (1 unit in US)
LocationBrea, CA, United States

Official product description

Access 2 Immunoassay System, Remanufactured, Catalog No. 386220 is an invitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analytes concentrations found in human body fluids.

Why the product was recalled

Due to the lack of updated thermal components, these instruments may be more susceptible to the internal case temperature exceeding the 18 to 36 degrees Celsius operating range, causing SYS flags. The instrument will not report results until the internal case temperature returns to the operating range. This situation may cause a delay in the reporting of patient results.

Distribution information

US, Morocco, Venezuela

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.