FDA recall record

Beckman Coulter Access 2 Immunoassay Analyzer, Catalog Number 81600N recall — FDA Z-1386-2025

Beckman Coulter identified that an issue for Access 2 Sample Motors (B48016) due to improper alignment, causing the motors to rub against the sample…

Class IIclassification
Z-1386-2025official record ID
Mar 26, 2025report date

Recall record at a glance

Official recordZ-1386-2025
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmBeckman Coulter, Inc.
Report dateMar 26, 2025
Recall initiation dateFeb 5, 2025
Quantity873 units
LocationChaska, MN, United States

Official product description

Access 2 Immunoassay Analyzer, Catalog Number 81600N

Why the product was recalled

Beckman Coulter identified that an issue for Access 2 Sample Motors (B48016) due to improper alignment, causing the motors to rub against the sample wheelbase plate and index sensor. The issue was traced to a worn tool used in production (at the supplier), affecting Access 2 sample motors. This could lead to a potential delay in reporting patient test results.

Distribution information

Worldwide - US Nationwide distribution.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.