Recall record at a glance
| Official record | Z-1065-2019 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Accelerate Diagnostics Inc. |
| Report date | Apr 3, 2019 |
| Recall initiation date | Mar 13, 2017 |
| Quantity | 81 |
| Location | Tucson, AZ, United States |
Official product description
Accelerate Pheno system, with Accelerate Pheno System Software 1.2.1 - 1.3.1, PSW000012B Product Usage: The Accelerate Pheno system is intended to measure signal intensity of fluorescent probes bound to nucleic acid in target and non-target organisms and to take time-lapse dark-field images of immobilized growing bacterial cells when used with Accelerate PhenoTest kits.
Why the product was recalled
Rare isolates of Enterobacteriaceae may generate a susceptible meropenem Minimum Inhibitory Concentration (MIC) using the affected device and a resistant result by broth microdilution. In order to mitigate any risk associated with these rare isolates, the firm has implemented an Expert Rule that suppresses the meropenem result for isolates that display a profile consistent with those that generated a false-susceptible result. The firm has responded to this risk by implementing a set of suppression rules, in a software update, to prevent incorrect results.
Distribution information
Worldwide Distribution - US Nationwide North America and Middle East US: TX, IL, TN, GA, CA, AR, MD, VA, Washington DC, IA, PA, NY, WA, and MO. OUS (Foreign): UK, Spain, Italy, and Germany
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Telephone.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.