FDA recall record

Hummingbird Med ABG-HM-1 Hummi Micro Draw Blood Transfer Device Product Usage: It is used recall — FDA Z-1636-2016

Hummingbird Med Devices, Inc. is recalling ABG-HM-1 Hummi Micro Draw Blood Transfer because the connection between the "Hummi" Y connector and the yellow…

Class Iclassification
Z-1636-2016official record ID
Jun 1, 2016report date

Recall record at a glance

Official recordZ-1636-2016
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusTerminated
Recalling firmHummingbird Med
Report dateJun 1, 2016
Recall initiation dateOct 1, 2015
Quantity11500 units
LocationLake Forest, CA, United States

Official product description

ABG-HM-1 Hummi Micro Draw Blood Transfer Device Product Usage: It is used for blood transfer and collection from Peripheral Arterial Line Catheters.

Why the product was recalled

Hummingbird Med Devices, Inc. is recalling ABG-HM-1 Hummi Micro Draw Blood Transfer because the connection between the "Hummi" Y connector and the yellow cannula hub may not be sufficiently secure enough and could separate while in use.

Distribution information

Distributed in the states of CA, KY, MD, GA, and IL.

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.