FDA recall record

Abbott MitraClip XTR Clip Delivery System, UDI: 08717648226366, Part: CDS0601-XTR, Sterile recall — FDA Z-1687-2019

Reports of implantable mitral valve repair system clips unexpectedly opening and becoming nonfunctional, resulting from unintended excessive force applied…

Class IIclassification
Z-1687-2019official record ID
Jun 5, 2019report date

Recall record at a glance

Official recordZ-1687-2019
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmAbbott Vascular
Report dateJun 5, 2019
Recall initiation dateMay 1, 2019
QuantityN/A
LocationTemecula, CA, United States

Official product description

Abbott MitraClip XTR Clip Delivery System, UDI: 08717648226366, Part: CDS0601-XTR, Sterile EO, Rx Only

Why the product was recalled

Reports of implantable mitral valve repair system clips unexpectedly opening and becoming nonfunctional, resulting from unintended excessive force applied during the Clip implantation. Excessive force can also result in unexpected movement of clip arms. The inability to close and remove the device has lead to surgery and additional intervention.

Distribution information

U.S.: FL, UT, IN, OK, TX, CO, LA, AZ, MO, AL, GA, CA, AR, KS, IL, HI, IA, KY, MA, DC, NY, NV, NM, NJ, MT, MD, NC, CT, MI, MN, OH, ND, NH, OR, PA, WI, WA, VA, WV, TN, SC, RI, DE, ME, NE, SD, MS, ID, VT. Foreign (OUS): AUSTRALIA, AUSTRIA, BELGIUM, BULGARIA, Canada, CROATIA, CZECH REPUBLIC, DENMARK, FINLAND, FRANCE, GERMANY, GREECE, HONG KONG, HUNGARY, INDONESIA, IRELAND, ISRAEL, ITALY, KUWAIT, LEBANON, MACAU, NETHERLANDS, NEW ZEALAND, NORWAY, POLAND, PORTUGAL, ROMANIA, RUSSIA, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SPAIN, SWEDEN, SWITZERLAND, THAILAND, TURKEY, UNITED KINGDOM

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.