Recall record at a glance
| Official record | Z-0881-2021 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Abbott Point Of Care Inc. |
| Report date | Jan 27, 2021 |
| Recall initiation date | Dec 22, 2020 |
| Quantity | 1,521,800 cartridges US and 2,641,053 cartridges OUS |
| Location | Princeton, NJ, United States |
Official product description
Abbott i-STAT EG7+ cartridge-IVD. Intended for the quantitative measurement of sodium, potassium, ionized calcium, pH, pCO2, pO2 and hematocrit in venous, arterial or capillary whole blood. List Number: 03P76-25
Why the product was recalled
Ionized calcium test on some i-STAT CG8+ and EG7+ cartridges may exhibit higher than expected reported results when stored at room temperature (18 to 30C) for periods of time in excess of 3 days for lot numbers *20100 to *20339 and 7 days for lot numbers *20340 and above Where * = lot letter A, K, L, M, N, W or Y.
Distribution information
Nationwide
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.