FDA recall record

Olympus Corporation of the Americas A20975A "Working insert, with ramp, one way" recall — FDA Z-0696-2018

Olympus has received complaints about fragments of adhesive which detached from inside the working channel of the referenced products. Cracking, chipping,…

Class IIclassification
Z-0696-2018official record ID
Mar 7, 2018report date

Recall record at a glance

Official recordZ-0696-2018
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmOlympus Corporation of the Americas
Report dateMar 7, 2018
Recall initiation dateAug 18, 2017
Quantity8741
LocationCenter Valley, PA, United States

Official product description

A20975A "Working insert, with ramp, one way" (Lots 13ZW to 172W; A20976A "Bridge, one way" (Lots 146W to 172W) and A20977A "Bridge, two way" (Lots 146W to 174W) Product Usage: A20975A: Working insert for endoscopic diagnosis and treatment in urologic applications. A20976A: Bridge for endoscopic diagnosis and treatment in urologic applications. A20977A: Bridge for endoscopic diagnosis and treatment in urologic applications.

Why the product was recalled

Olympus has received complaints about fragments of adhesive which detached from inside the working channel of the referenced products. Cracking, chipping, missing pieces, and delamination of the adhesive have also been reported. In one case, a fragment of the adhesive was detected inside a patient's urethra during a procedure and was retrieved without injury to the patient. The manufacturer, Olympus Winter & lbe GmbH ("OWi") has determined that adhesive can be dislodged during the intended use of the cystoscopy bridge or working insert, e.g., when inserting an instrument through the working channel. As a result, a fragment of the adhesive may fall inside the patient's bladder, urethra or uterus and would need to be retrieved, extending the length of the procedure or requiring additional surgical treatment. The adhesive is used to seal gaps inside the working channel of the cystoscopy bridges and working inserts. In 2013, the adhesive used until then was discontinued and had to be replaced. Current investigations revealed that the durability of the new adhesive is insufficient

Distribution information

US Nationwide Distribution

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records