FDA recall record

Baxter Healthcare A) Product Code 2N8371: Non-DEHP Y-Type Catheter Extension Set Kit, 6.0" recall — FDA Z-1271-2013

Customers have reported that when separating an individual package from its attached grouping, the adjacent package has opened, compromising its sterile…

Class IIclassification
Z-1271-2013official record ID
May 15, 2013report date

Recall record at a glance

Official recordZ-1271-2013
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmBaxter Healthcare Corp.
Report dateMay 15, 2013
Recall initiation dateMar 28, 2013
QuantityA) Product Code 2N8371: 513,000 units
LocationDeerfield, IL, United States

Official product description

A) Product Code 2N8371: Non-DEHP Y-Type Catheter Extension Set Kit, 6.0", Vol. 1.0 mL, 2 Luer Activated Valves, Male Luer Lock Adapter, Single use only Product Usage: The IV extension tubing kits are intended to provide users with a ready to use set up for the administration and withdrawal of fluids avoiding the need to acquire/assemble components from various manufacturers. The IV extension tubing sets are indicated for use with a vascular access device for administration and withdrawal of fluids.

Why the product was recalled

Customers have reported that when separating an individual package from its attached grouping, the adjacent package has opened, compromising its sterile barrier properties.

Distribution information

Worldwide Distribution - USA Nationwide and the countries of Puerto Rico, Brazil, Canada and Columbia.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.