Recall record at a glance
| Official record | Z-1270-2013 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Baxter Healthcare Corp. |
| Report date | May 15, 2013 |
| Recall initiation date | Mar 28, 2013 |
| Quantity | A) Code 2N8220: 5,950; B) Code 2N8221: 6,150; C) Code 2N8222: 1,850; D) Code 2N8223: 7,900; E) Code 6N8220: 52,200; F) Code 6N8222: 1,200 |
| Location | Deerfield, IL, United States |
Official product description
A) Product Code 2N8220: Non-DEHP Catheter Extension Set Kit, Single use only, includes: 1) 7.7", Vol. 1.10 mL, Luer Activated Valve for IV Access, Male Luer Lock Adapter, 2) Medium Catheter Stabilization Device, and 3) Foam Tape Strips; B) Product Code 2N8221: Non-DEHP Catheter Extension Set Kit, Single use only, includes: 1) 7.1", Vol. 0.85 mL, Injection Site Male Luer Lock Adapter with Retractable Collar, 2) Medium Catheter Stabilization Device, and 3) Foam Tape Strips; C) Product Code 2N8222: Non-DEHP Catheter Extension Set Kit, Single use only, includes: 1) 8.2", Vol. 0.57 mL, Luer Activated Valve for IV Access, Male Luer Lock Adapter, 2) Medium Catheter Stabilization Device, and 3) Foam Tape Strips; D) Product Code 2N8223: Non-DEHP Catheter Extension Set Kit, Single use only, includes: 1) 7.7", Vol. 0.40 mL, Injection Site Male Luer Lock Adapter, 2) Medium Catheter Stabilization Device, and 3) Foam Tape Strips; E) Product Code 6N8220: V-LINK Luer Activated Device with VITALSHIELD Protective Coating Non-DEHP Catheter Extension Set Kit, Single use only, includes: 1) 7.6", Vol. 1.0 mL, Catheter Extension Set, Luer Activated Device with Silver for IV Access, Male Luer Lock Adapter with Retractable Collar, 2) Medium Catheter Stabilization Device, and 3) Foam Tape Strips; F) Product Code 6N8222: V-LINK Luer Activated Device with VITALSHIELD Protective Coating Non-DEHP Catheter Extension Set Kit, Single use only, includes: 1) 8.2", Vol. 0.5 mL, Catheter Extension Set, Luer Activated Device with Silver for IV Access, Male Luer Lock Adapter, 2) Medium Catheter Stabilization Device, and 3) Foam Tape Strips Product Usage: The IV extension tubing kits are intended to provide users with a ready to use set up for the administration and withdrawal of fluids avoiding the need to acquire/assemble components from various manufacturers. The IV extension tubing sets are indicated for use with a vascular access device for administration and withdrawal of fluids.
Why the product was recalled
Customers have reported that when separating an individual package from its attached grouping, the adjacent package has opened, compromising its sterile barrier properties.
Distribution information
Worldwide Distribution - USA Nationwide and the countries of Puerto Rico, Brazil, Canada and Columbia.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.