Recall record at a glance
| Official record | Z-1268-2013 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Baxter Healthcare Corp. |
| Report date | May 15, 2013 |
| Recall initiation date | Mar 28, 2013 |
| Quantity | A) Product Code 2N1191: 53,850; B) Product Code 2N1206: 101,900 units |
| Location | Deerfield, IL, United States |
Official product description
A) Product Code 2N1191: Non-DEHP Y-Type Catheter Extension Set, 4.8", Vol. 0.6 mL, Male Luer Lock Adapter, Single use only; B) Product Code 2N1206: Non-DEHP I.V. Catheter Extension Set, 5.7", Vol. 0.7 mL, Male Luer Lock Adapter with Retractable Collar, Single use only Product Usage: These devices are indicated for use with a vascular access device for administration and withdrawal of fluids.
Why the product was recalled
Customers have reported that when separating an individual package from its attached grouping, the adjacent package has opened, compromising its sterile barrier properties.
Distribution information
Worldwide Distribution - USA Nationwide and the countries of Puerto Rico, Brazil, Canada and Columbia.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.