FDA recall record

C.R. Bard, Inc., Urological Division 8 Fr. Bardex Pediatric All-Silicone Foley Catheter with recall — FDA Z-0010-2013

The 8 Fr. Bardex All-Silicone Foley Catheter with 3cc balloon had an incorrect inflation volume (10 ml instead of 5 ml) printed on the valve cap.

Class IIclassification
Z-0010-2013official record ID
Oct 10, 2012report date

Recall record at a glance

Official recordZ-0010-2013
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmC.R. Bard, Inc., Urological Division
Report dateOct 10, 2012
Recall initiation dateSep 7, 2012
Quantity4,032 units
LocationCovington, GA, United States

Official product description

8 Fr. Bardex Pediatric All-Silicone Foley Catheter with a 3cc balloon. The product is packaged individually with 12 units per corrugated case. The product is sold sterile for single use and is used to provide continuous urinary drainage from the bladder.

Why the product was recalled

The 8 Fr. Bardex All-Silicone Foley Catheter with 3cc balloon had an incorrect inflation volume (10 ml instead of 5 ml) printed on the valve cap.

Distribution information

Nationwide distribution: USA and Puerto Rico

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.